Measurement Functions | AC/DC Withstanding Voltage & IR |
Interface | USB-HOST 、RS232/RS485 、HANDLER 、ALARM、REMOTE |
Standard Configuration | 5kV、100mA、500VA |
Comparator | UPPER/LOWER |
Protect Function | OV/OC |
Other Functions | LIST Edit Discharge Function 5“ LCD + Touch panel |
AC Withstand Voltage | ● |
Rate Ouput | AC 5.000kV/100mA |
Output Power | 500VA |
Output Voltage | 100~5000 V AC Resolution: 1V |
Voltage Accuracy | 2% of reading + 5 digits |
Output Regulation | 2% of reading +5V |
Voltage Output Stability | ±(0.4% of setting +1V)/ min,From no load to full load |
Output Ripple | Resistive Load:<2% |
Output Frequency | 50Hz/60Hz |
Short-circuit Current | >200mA |
Voltage Measurement | (0.10 ~ 5.00)kV,Resolution: 0.01kV |
Voltage Accuracy | (1.5% of reading+1 counts) |
Measurement Current | 0.000~3.500/3.00~100.00 mA , Resolution:0.001/0.01 mA |
Current Accuracy | 2% of reading + 5 digits |
DC Withstand Voltage | ● |
Rate Output | DC 6kV/10mA |
Output Power | 60VA, ≥90% |
Output Voltage | 0.1kV-6.000kV DC Resolution: 1V |
Voltage Accuracy | 2% of reading + 5 digits |
Output Regulation | ±(2% of setting +5V),From no load to full load |
Output Ripple | <5%(6kV/1mA at Resistive Load) |
Current Measurement | 0.0~350.0/300~3500/3000~9999 μA/10mA Resolution: 0.1/1/10μA/0.01mA |
Current Accuracy | 2% of reading +5 Counts |
Insulation Resistance | ● |
Rate Output | DC 2.5kV/9999MΩ |
Output Voltage | 100~2500V DC Resolution: 1V |
Voltage Accuracy | 2%of setting +5V |
Drop Voltage | Not less than rated voltage 90%,drop resistance 10MΩ(1% of range) |
Measurement Range | 0.1MΩ-10GΩ |
IR Accuracy | Range:(1 ~9999) MΩ,Resolution: 1MΩ Accuracy:(100-499) V, 1~1000,±(8% of reading +2counts) 1000~2000,±(12% of reading + 2counts) (500-2500V), 1~199,±(2% of reading + 2counts), 200~999, ± (5% of reading + 2counts) 1000~9999, ± (15% of reading + 2 counts) |
Power Supply | 198V~242V AC 50/60Hz |
Dimension & Weight | 400(W)x200(H)x503.5(D) 20kg |